Models 140, 141, and 142 Conductivity Sensors Instruction Sheet PN 51A-140/rev .E December 2010 SENSOR SPECIFICA TIONS For additional information, please visit our website at www .
Keep 1/4 inch (6 mm) clearance between electrodes and piping. The electrodes must be completely sub- merged in the process liquid, i.e., to the level of the threaded connection. See Figure 1 for recommended orientation. If the sensor is installed in a side stream with the sample draining to open atmosphere, bubbles may accumulate on the electrodes.
3 MODEL 140,141, and 142 sensors INST ALLA TION FIGURE 3. Inst alling model 141 sensor DWG . NO. REV . 40014204 B INST ALLA TION – 141 SENSOR 1. Install the sensor in a 3/4-inch NPT weldalet or in a 1-inch pipe tee. 2. Remove the plastic shipping cap from the sensor .
RTD RTD in RTD sense RTD return MODEL 140,141, and 142 sensors WIRING WIRING All 140 series sensors have a junction box mounted on the back of the sensor . Wiring connections in the junction box are shown in Figure 5. Figure 5. Sensor junction box wiring.
5 MODEL 140,141, and 142 sensors WIRING FIGURE 8. Model Xmt-C-10 Wiring (Panel) FIGURE 9. Model Xmt-C-1 1 Wiring (Pipe or Wall) RCV B RCV A RSHLD DRVB RTN SENSE RTD IN SHLD CLEAR WHITE/RED RED DRV A DSHLD WHITE CLEAR GRA Y ORANGE CLEAR TB1 TB2 FIGURE 10.
RETRACTING AND INSERTING THE MODEL 140 SENSOR RETRACTING THE SENSOR 1. Push in on the sensor junction box and slowly loosen the sensor compression fitting nut by reversing the sensor tightening procedure illustrated in Figure 2.
MODEL 140,141, and 142 sensors WIRING 7 3. Slowly open the valve. 4. Insert the sensor up to the desired insertion depth and turn the sensor compression fitting nut until it is finger tight. 5. Position the entire sensor for easy access to the ball valve handle, sensor compression fitting nut and J-box terminal block.
MODEL 140,141, and 142 sensors CALIBRA TION 8 CALIBRA TION Model 140 sensors are not calibrated at the factory . The cell const ant on the label is a nominal value only .
9 MODEL 140,141, and 142 sensors MAINTENANCE MAINTENANCE CLEANING THE SENSOR Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.
10 MODEL 140,141, and 142 sensors ACCESSORIES PN Description 23724-00 Ball valve kit for Model 140 sensor 23730-00 Process compression fitting kit (3/4-inch NPT) for Model 140 sensor 23731-00 Process .
11 TROUBLESHOOTING PROBLEM PROBABLE CAUSE SOLUTION Off-scale reading Wiring is wrong. V erify wiring. T emperature element is open or shorted. Check temperature element for open or short circuits. See Figure 14. Sensor is not in the process stream. Be sure the sensor is completely submerged in the process stream.
Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. 8 Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .
デバイスEmerson 140の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Emerson 140をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはEmerson 140の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Emerson 140の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Emerson 140で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Emerson 140を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はEmerson 140の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Emerson 140に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちEmerson 140デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。