Model 228 Submersion/Inser tion T oroidal Sensor Instruction Sheet PN 51A-228/rev .P December 2010 SPECIFICA TIONS - SENSORS MODEL Wetted Materials T emperature Pressure CRN pressure* 228-02 PEEK (gla.
MODEL 228 INST ALLA TION INST ALLA TION Keep at least 1 in. (2.5 cm) between the sensor and pipe walls. If the clearance is too small, calibrate the sensor in place. Ensure that the sensor is completely submerged in liquid. Mounting the sensor in a vertical pipe run with the flow from bottom to top is best.
3 MODEL 228 WIRING FIGURE 6. Wiring 228-54 and 228-56 sensors to Model 54eC analyzer FIGURE 5. Wiring 228-54 and 228-56 sensors to Model 1056 and 56 analyzers.
MODEL 228 WIRING FIGURE 7. Wiring 228-54 and 228-56 sensors to Model Xmt-T p anel mount transmitter FIGURE 8. Wiring 228-54 and 228-56 sensors to Model Xmt-T pipe/wall mount transmitter 4.
5 MODEL 228 WIRING FIGURE 9. Wiring 228-54 and 228-56 sensors to Model 1066 transmitter FIGURE 10. Wiring 228-54 and 228-56 sensors to Model 5081-T transmitter RCV B RCV A RSHLD DRV B 228-54 RTN SENSE.
MODEL 228 WIRING FIGURE1 1. Wiring sensors through a remote junction box Wire sensors point to point. For wiring at the analyzer end, refer to the appropriate analyzer wiring diagram. For interconnecting cable 23294-00, use the 2265-54 wiring diagram.
7 MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES INSERTION - RETRACTION ASSEMBLIES MANUAL RETRACTION ASSEMBL Y (PN 2331 1-01) INST ALLA TION 1. Loosen the collet nut and retract the sensor tube into the retraction chamber . See Figure 14. 2. Loosen the union nut and separate the retraction chamber from the assembly .
MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES FIGURE 14. Manual Retraction Assembly MILLIMETER INCH RETRACTION 1. Make certain that the system pressure is less than 35 psig (342 kPa abs). Push in on the sensor using the top of the junction box. Slowly loosen the collet nut.
9 MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES INST ALLA TION 1. Thread the sensor cable through the tube into the junction box. Screw the sensor into the tube. Hand tighten the sensor an additional half turn once the gas- ket is seated. See Figure 15.
MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES 1. Retract the sensor into the retraction chamber and fully close the ball valve. 2. Drain the retraction chamber content s using the 1/8 in. flush ports. 3. (Mechanical) Mark the location of the nut housing cap and retraction collar on the sensor tube.
11 MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES FIGURE 16. Mechanical Retraction Assembly Replacement Part s.
FIGURE 17. Manual Retraction Assembly Replacement Part s MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES 12.
13 MODEL 228 CALIBRA TION CALIBRA TION The nominal cell constant of the Model 228 sensor is 2.7/cm. The error in the cell constant is about ± 10%, so con- ductivity readings made using the nominal cell constant will have an error of at least ± 10%. W all ef fects, discussed below , will likely make the error greater .
MAINTENANCE Generally , the only maintenance required is to keep the opening of the sensor clear of deposits. Cleaning frequency is best determined by experience. MODEL 228 MAINTENANCE W ARNING Retraction chamber contents may be under pressure. CAUTION Be sure sensor has been cleaned of process liquid before handling.
15 FIGURE 19. Resist ance check. Disconnect leads from analyzer before measuring resist ances. T ABLE 1. T emperature Resist ance 10 ° C 103.9 Ω 20 ° C 107.
Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .
デバイスEmerson 228の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Emerson 228をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはEmerson 228の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Emerson 228の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Emerson 228で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Emerson 228を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はEmerson 228の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Emerson 228に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちEmerson 228デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。