Page 1 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Leica HER2 FISH System - 30 T est Instructions For Use For use on Leica Biosystems’ BOND-MAX and BOND-III System.
English Page 2 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Contents Intended Use .................................................................................................
Page 3 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Intended Use For in vitro diagnostic use The Leica HER2 FISH System .
English Page 4 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 The characteristic morphology of the tissue and the spatial distribution of oncogene copies in individual uncultured primary breast carcinomas are retained.
Page 5 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English The Leica HER2 FISH System - 30 T est is for use only on the Leica BOND-MAX and BOND-III System.
English Page 6 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 T issue sections mounted on charged slides (Leica BOND Plus Slides S21.21 13) can be held for up to 12 months at 2–8 °C before staining.
Page 7 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English bandpass uorescence microscope lter sets optimized for use with the Leica HER2 FISH System - 30 T est are available for most microscope models.
English Page 8 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 5. T urn on the PC attached to the Leica BOND-MAX and BOND-III System.
Page 9 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Signal Assessment and Enumeration T o assess the signal quality and enumerate the HER2 and CEP17 signals follow the process below: 1.
English Page 10 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Recommended Method for LSI HER2 to CEP17 Ratio Determination T o determine the LSI HER2 to CEP17 ratio, use the following method: 1.
Page 1 1 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Figure 1: Interpretation Guide Leica HER2 FISH S ystem - 30 T est Interpretation Guide Count as 2 orange signals and 1 green signal Do not count.
English Page 12 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Leica HER2 FISH S ystem - 30 T est Score Sheet 20 Nuclei Signal Cou.
Page 13 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Quality Control Use of Control Slides It is recommended that a Leica HER2 FISH Control Slide is included in each test run to monitor assay performance and to assess the accuracy of signal enumeration.
English Page 14 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Unexpected FISH staining, or variations in the staining, may be a result of alterations in the expression levels of the encoding genes.
Page 15 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English 2x2 Concordance Results Leica BOND-MAX System Data was grouped as negative (<2.
English Page 16 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 2x2 Concordance Results Leica BOND-III System Data was grouped as negative (<2.
Page 17 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Precision T esting – Leica BOND-MAX System A. Within Run Precision Study The within run precision study was performed in a randomized and blinded fashion.
English Page 18 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 F . Lot-to-Lot Precision Study The lot-to-lot precision study was performed in a randomized and blinded fashion.
Page 19 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English K. Between Observer Precision Study The between observer precision study was performed in a randomized and blinded fashion.
English Page 20 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 denaturation time is recommended.
Page 21 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English T roubleshooting Problem Probable C ause Remedial Action No or weak.
English Page 22 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Problem Probable C ause Remedial Action Nonspecic background uorescent signal/staining Inadequate post hybridization washing Increase post hybridization wash incubation time.
Page 23 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English References 1. Bargmann CI, Hung MC, Weinberg RA. The neu oncogene encodes an epidermal growth factor related protein. Nature 1986;319:226–30.
Page 24 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 License Agreement This product contains PathV ysion FISH probes supplied by Abbott Molecular Inc. PathV ysion, LSI and CEP are a trademark of Abbott Molecular Inc.
デバイスLeica TA9217の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Leica TA9217をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはLeica TA9217の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Leica TA9217の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Leica TA9217で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Leica TA9217を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はLeica TA9217の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Leica TA9217に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちLeica TA9217デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。