National Instrumentsメーカー9211Aの使用説明書/サービス説明書
ページ先へ移動 of 19
USER GUIDE AND SPECIFICA TIONS USB-9211/9211A 4-Channel, 24-Bit Thermocouple Input Devices This user guide describes how to use the National Instruments USB-9211/9211A devices and lists the specifications.
USB-9211/9211A User Guid e and Specifications 2 ni.com Dimensions Figure 2 shows the USB-9211/9211A devi ce dimensions. Figure 2. USB-9211/9211A Series Devices in Inches (Millimeters) Safety Guidelines Operate the USB-9211/9211A only as described in these operating instructions.
© National Instruments Cor poration 3 USB-9211/9211A User Guide and Specifications Safety Guidelines for Hazardous Vo ltages If hazardous voltages are connected to the module, take the following precautions. A hazardous vo ltage is a voltage greater than 42.
USB-9211/9211A User Guid e and Specifications 4 ni.com Related Documentation If your device uses NI-D A Qmx Base, refer to ni.com to do wnload the software. Installing the USB-9211/9211A Device Installing the Software Before installing the device, you must install the software you plan to use with the device.
© National Instruments Cor poration 5 USB-9211/9211A User Guide and Specifications 3. Align the I/O module with the carrier, as sho wn in Figure 3. Figure 3. Module Installation 4. Squeeze the latches and insert the NI 9211 module into the carrier . 5.
USB-9211/9211A User Guid e and Specifications 6 ni.com Mounting the USB-9211 /9211A to a Panel Threaded inserts are located in the USB-921 1/9211A for mounting it to a panel.
© National Instruments Cor poration 7 USB-9211/9211A User Guide and Specifications Wiring the USB-9211/9211A The USB-9211/9211 A has a 10-terminal, detachabl e high voltage screw terminal enclosure that provides connections for four therm ocouple input channels.
USB-9211/9211A User Guid e and Specifications 8 ni.com Refer to T able 3 for the termin al assignments for each channel. Assembling the High V oltage Screw T erminal Backshell The high voltage screw terminal back shell must be instal led when using hazardous voltages (>42.
© National Instruments Cor poration 9 USB-9211/9211A User Guide and Specifications Figure 7. High V oltage Screw T erminal Backsh ell USB-9211/9211A Circuitr y The USB-9211/9211A channels share a common gro und that is isolated from the chassis and the host comput er.
USB-9211/9211A User Guide and Specifications 10 ni.com Figure 8. Input Cir cuitry for One Channel If thermocouples are connected to the USB-921 1/9211A, the gain and of fset errors resulting from the so ur ce impedance of the thermocouples are negligible for most applications.
© National Instruments Corporation 11 USB- 9211/9211A User Gu ide and Specifications Measurement Accuracy fo r the Different T ypes of Thermocouples Figures 9, 10, 11, 12, and 13 show the typical and maximum errors for the different thermocouple types when used with the USB-9211/9211A over the full temperature range.
USB-9211/9211A User Guide and Specifications 12 ni.com Figure 10. T ype E and T Errors Figure 11. T ype J and N Errors –200 0 200 400 600 800 1000 0 2 3 4 Max ov er T emp Range with A utozero Max at.
© National Instruments Corporation 13 USB- 9211/9211A User Gu ide and Specifications Figur e 12. T ype K Errors Figure 13. T ype R and S Errors Cold-Junction T emperatur e Measurement Accuracy Heat f.
USB-9211/9211A User Guide and Specifications 14 ni.com Minimizing Thermal Gradients Thermocouple wire can be a significan t source of thermal gradients if it conducts heat or cold directly to terminal j unctions. To min imize these errors, follow these guidelines: • Use small-gauge thermocouple wire.
© National Instruments Corporation 15 USB- 9211/9211A User Gu ide and Specifications Cold-junction compensation sensor accuracy 0 to 60 °C ...................... .................. 0.6 °C (1.1 °F) typ, 1.3 °C (2.3 °F) max Conversion time .......
USB-9211/9211A User Guide and Specifications 16 ni.com Offset error from source impedance ...... . 0.05 µ Vt y p ,0 . 0 7 µ Vm a x p e r Ω source impedance due to input current Power Requirements Current consumption from USB .............500 mA, max Suspend mode .
© National Instruments Corporation 17 USB- 9211/9211A User Gu ide and Specifications V oltages Connect only voltages that are within these limits. Channel-to-COM .
USB-9211/9211A User Guide and Specifications 18 ni.com Maximum altitude ................. ..................2,000 m (at 25 °C ambient temperature) Pollution Degree (IEC 60664) .... ............2 Electromagnetic Compatibility Emissions ...............
National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for m ore inform ation about Natio nal Instruments trademarks. Other product and compan y names mentioned herein are t rademarks or trade names of their respective companie s.
デバイスNational Instruments 9211Aの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
National Instruments 9211Aをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはNational Instruments 9211Aの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。National Instruments 9211Aの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。National Instruments 9211Aで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
National Instruments 9211Aを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はNational Instruments 9211Aの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、National Instruments 9211Aに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちNational Instruments 9211Aデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。