National InstrumentsメーカーSCC-TC02の使用説明書/サービス説明書
ページ先へ移動 of 20
USER GUIDE SCC-TC Series Thermocouple Input Modules The SCC-TC Series thermocouple input modules, SC C-TC01 and SCC-TC02, accept input signals from B-, E-, J-, K-, N-, R-, S-, and T-type thermocouples. The thermocouple inpu ts are filtered and passed into a differential amplif ier with a gain of 100.
SCC-TC Series Thermocouple Input Modules User Guide 2 ni.com When symbol is marked on a product, it denotes a warning advising you to take precautions to av oid electrical shock. When symbol is marked on a product, it denotes a component that may be hot.
© National Instruments Corpor ation 3 SCC-TC Series Th ermocouple Input Mo dules User Guide ❑ Numbers 1 and 2 Phillips screwdri vers ❑ W i re insulation stripper s ❑ NI-D A Q (current version) for Wi ndows 2000/NT/XP Note Software scaling of measurements is not supported on the Macintosh operating system.
SCC-TC Series Thermocouple Input Modules User Guide 4 ni.com A blue stripe on the label identifies the SCC-TC0 X as an analog input module. The label also displays the ico n sho wn in Figure 1. Figure 1. SCC-TC0 X Icon Y ou can plug the SCC-TC0 X into any analog input so cket on the SC-2345.
© National Instruments Corpor ation 5 SCC-TC Series Th ermocouple Input Mo dules User Guide The SCC-TC02 has a f ixed scre w-term inal receptacle and a remov able screw-terminal block, as sho wn in Figure 2.
SCC-TC Series Thermocouple Input Modules User Guide 6 ni.com The amplified thermocouple signal and cold-junction sensor sig nal are measured by E Serie s D A Q device channel X and channel X + 8 respecti vely , where X is 0 to 7 depending on where you plug the SCC-TC0 X .
© National Instruments Corpor ation 7 SCC-TC Series Th ermocouple Input Mo dules User Guide Y ou can use the follo wing formulas to con vert the cold-junction sensor voltage to cold-junction temperature: where T K is the temperature in kelvin. where a = 1.
SCC-TC Series Thermocouple Input Modules User Guide 8 ni.com Detecting Open Thermocouples The SCC-TC0 X contains a 10 M Ω pull-up resistor that connects to + 5 V to detect open thermocouples. To determine if you have an open thermocouple, check whether the corresponding E Series DAQ dev ice channel is saturated.
© National Instruments Corpor ation 9 SCC-TC Series Th ermocouple Input Mo dules User Guide NI-DAQmx Complete the following steps to configure th e SCC-TC module using NI-DAQmx. 1. Open Measurement & A utom ation Explorer . 2. Right-click Devices and Interfaces and select Create New .
SCC-TC Series Thermocouple Input Modules User Guide 10 ni.com 6. Under Accessory: , select SC-23 45 . 7. Click the Configure button. A ne w window appears listing the sock ets (connector reference designators) of the SC-2345. 8. Select the SC-2345 socket where you installed an SCC.
© National Instruments Corpor ation 11 SCC-TC Series Thermocouple Input Modules User Guide 8. Enter the specif ic values for your application in the Setti ngs tab . Context help information for each setting is pro vided on the right side of the screen.
SCC-TC Series Thermocouple Input Modules User Guide 12 ni.com Calibrating the SCC-TC0 X Using a Thermocouple Calibrator To calibrate the system using a thermocouple calibrator, complete the following steps: 1. Make sure that the thermo couple connections at both the SCC and the thermocouple calibrator ar e at the same temperature.
© National Instruments Corpor ation 13 SCC-TC Series Thermocouple Input Modules User Guide where V TC is thermocouple voltage. V ESERIES is E Series D A Q device v oltage. This step provides proper scaling for the th ermocouple amplifier in the SCC-TC0 X .
SCC-TC Series Thermocouple Input Modules User Guide 14 ni.com Amplifier Characteristics Open thermocouple detectio n current.....250 nA max 1 (assuming floating ther mocouple) Common-mode rejection rati o ................110 dB min 1 Output range ....
© National Instruments Corpor ation 15 SCC-TC Series Thermocouple Input Modules User Guide Cold-Junction Sensor Cold-junction sensor accuracy .............. . ± 0.4 ° C max from 15 to 35 ° C, ± 0.75 ° C max from 0 to 15 ° C and 35 to 55 ° C Output.
SCC-TC Series Thermocouple Input Modules User Guide 16 ni.com Physical TC01 dimensions ........................... .........7.8 cm by 2.92 cm by 1.85 cm (3.1 in. by 1.15 in. by 0.73 in.) TC02 dimensions ........................... .........8.89 cm by 2.
© National Instruments Corpor ation 17 SCC-TC Series Thermocouple Input Modules User Guide Note For UL and other safety certifications, refer to the product label, or visit ni.com/hardref.nsf , search by model number or product line, and click the appropriate link in the Certification column.
SCC-TC Series Thermocouple Input Modules User Guide 18 ni.com I/O Connector Pin Assignments Figure 4 shows the I/O connector pins on the bottom o f the module. Figure 4. SCC Module Bottom View T able 1 lists the signal connection corresponding to each pin.
© National Instruments Corpor ation 19 SCC-TC Series Thermocouple Input Modules User Guide 13 + 15 V 14 –15 V 15 — 16 — 17 — 18 — 19 — 20 — Table 1.
National Instruments™, NI™, ni.com ™, and NI-D AQ™ are trademarks of Nationa l Instruments Corporation. Product and com pany names mentioned herein are trademarks or trade names of their respective companies.
デバイスNational Instruments SCC-TC02の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
National Instruments SCC-TC02をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはNational Instruments SCC-TC02の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。National Instruments SCC-TC02の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。National Instruments SCC-TC02で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
National Instruments SCC-TC02を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はNational Instruments SCC-TC02の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、National Instruments SCC-TC02に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちNational Instruments SCC-TC02デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。