Omega EngineeringメーカーHH504の使用説明書/サービス説明書
ページ先へ移動 of 15
OMEGA HH504 Digital Thermometer.
OMEGAnet SM On-Line Service http://www.omega.com Internet e-mail info@omega.com Servicing Nor th America: USA: ISO 9001 Certified Canada: One Omega Drive, Box 4047 976 Bergar Stamford, CT 06907-0047 Laval (Quebec) H7L5A1 Tel: (203) 359-1660 Tel: (514) 856-6928 FAX: (203)359-7700 FAX: (514) 856-6886 e-mail: info@omega.
1 WARNING To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC. WARNING To avoid damage or burns, do not make temperature measurement in microwave ovens. The symbol on the instrument indicates that the operator must refer to an explanation in this manual.
2 SPECIFICATIONS ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable. Measurement Range: Pt385(100 W ) -200°C to 800°C, (-328°F to 1472°F) Pt3926(100 W ) -200°C to 630°C, (-328°F to 1166°F) Resolution: 0.
3 ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 4½ digit liquid crystal display (LCD) with maximum reading of 19999. Overload: "----.
2 3 1 OPERATING INSTRUCTIONS Power Switch The key turns the thermometer on or off. In the data SET mode, can not power off, must leave data SET mode then power off. °C/°F Selecting the Temperature Scale (Main display) Reading the main displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F).
5 4 6 5 T1/ T2/T1-T2 Main display Input Selection The input selection indicates which input is selected for main display; T1 probe, T2 probe or the difference between the two probes (T1-T2), when the thermometer is turned on, it is set to T1.
6 7 The true average of all the reading taken over at least 22 hours preiod can be displayed. If 22 hours is exceeded, new averages are no longer calculated. The last calcalated value is retained as the average reading, but the actual minimum and maximum reading will continue to be captured.
8 9 °C/°F Selecting the Temperature Scale (Second display) Readings the second displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F). When the thermometer is turned on, it is set to the temperature scale that was in use when the thermometer was last turned off.
8 10 11 12 4. When the thermometer is turned on. The Relative set value and Hi/Lo Limits set value that was in use when thermometer was last turned off set values.
9 RTD PROBE CONNECTION P/N : 17650-0505 RW1=RW2=RW3 RW= WIRE RESISTANCE G + - RTD 100 W Pt385 (Pt3926) WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the rmometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery.
10 TEMPERATURE VS RESISTANCE TABLE(ITS90) °C Pt385 Pt3926 -200°C 18.521 W 16.996 W -100°C 60.256 W 59.479 W 0°C 100.000 W 100.000 W 100°C 138.505 W 139.272 W 200°C 175.856 W 177.362 W 300°C 212.052 W 214.275 W 400°C 247.092 W 250.018 W 500°C 280.
1 HH 504 / HH 505 CA LIBRA TION PROCEDURE Note: The following calibration procedure s hould perform only by qualified technicians who have access to the items as fol lowing. Equipment: The class of calibrator ha d better 10 times g reater than the measur ed meter.
WARRANTY / DISCLAIMER OMEGA ENGINEERING, INC . warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA W arranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! TEMPERATURE þ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies þ Wire: Thermoc.
デバイスOmega Engineering HH504の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Omega Engineering HH504をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはOmega Engineering HH504の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Omega Engineering HH504の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Omega Engineering HH504で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Omega Engineering HH504を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はOmega Engineering HH504の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Omega Engineering HH504に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちOmega Engineering HH504デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。