Shop-VacメーカーXPO100の使用説明書/サービス説明書
ページ先へ移動 of 20
Service Manual DEALER: Keep this m anual. The procedur es in this m anual MUST be performed by a qualifi ed technician. For mo re inform ation reg arding In va ca re p ro du cts , pa rts, and services, please v isit www.
XPO 2 ™ 2 Part No 1150748 WARNIN G DO NOT USE THIS PRODUCT OR ANY AVAIL ABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UNDERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIA L SUCH AS OWN ER’S MANUALS , SERVICE MANUALS OR INS TRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT.
TABL E OF CON TENT S Part No 1150748 3 XPO 2 ™ TABLE OF CONTENT S SPE CI AL NOT ES ..... ...... ...... ...... ....... ...... ...... ...... ....... ...... ...... ...... ...... . 4 TYP ICA L P RO DUC T P ARAM ET ERS ...... ...... ...... ...... .......
SPECIAL NOTES XPO 2 ™ 4 Part No 1150748 SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definiti ons of the signal words.
TYPICAL PRODUCT PARAMETERS Part No 1150748 5 XPO 2 ™ TYPICAL PRODUCT PARAMETERS Direct Current Ty pe B F e qui pmen t Attention - Consider Accompanying D ocuments DO NOT smoke Class II, Double Insul.
TYPICAL PRODUCT PARAMETERS XPO 2 ™ 6 Part No 1150748 Regul atory Lis ting Battery Duration (times are approximate): Setting 1 = 3.5 hours Setting 2 = 2.5 hours Setting 3 = 2.0 hours Setting 4 = 1.5 hours Setting 5 = 1.0 hours Battery Recharge Ti me: NOTE: Recharge time increases if battery is charging while unit is running.
SECTION 1—IMPORTANT SAFEGUARDS Part No 1150748 7 XPO 2 ™ SECTION 1—IMPORTANT SAFEGUARDS In order to ensure the safe in stallation, assembly and operation of the XPO 2 P ortable Concentrator these ins tructions MUST be follow ed.
SECTION 1—IMPORTANT SAFEGUARDS XPO 2 ™ 8 Part No 1150748 Radio F requency Interference This equi pment has been tested and found to comply with EMC limi ts specifi ed by IEC/ EN 60601-1-2. These limi ts are designed to provide a reasonable protection against electromagnetic interference in a typical medi cal in stallation.
SECTION 2—COMPONENT REPLACEMENT Part No 1150748 9 XPO 2 ™ SECTION 2—COMPONENT REPLACEMEN T CAUTION Ensure not to damage ribbon cable when removing or installing front cove r . DO NOT use Methyl Ethyl Ket one (MEK). This wi ll damage the plastic.
SECTION 2—COMPONENT REPLACEMENT XPO 2 ™ 10 Part No 1150748 1. Remov e label from the top of the XPO 2 (FIGURE 2.2). NOTE: This label adheres to the f ront and rear sections of the XPO 2 . This la bel MUST be rem oved to separate the two sections. FIGURE 2.
SECTION 2—COMPONENT REPLACEMENT Part No 1150748 11 XPO 2 ™ 4. While holding the unit tog ether rotate the XPO 2 and position it so that the back panel label is faci ng down. NOTE: The ribbon cable i s connected to the front cov er and the pc board.
SECTION 2—COMPONENT REPLACEMENT XPO 2 ™ 12 Part No 1150748 4. T ransfer the 5 / 8 -inch tubing to the intake of the new filter . NOTE: The flow arrow points away from the intake. 5. Connect the 11½- i nch tubi ng to the outlet of the i nlet fi lter (Detail “ A” of FIGURE 2.
SECTION 2—COMPONENT REPLACEMENT Part No 1150748 13 XPO 2 ™ NOTE: When installing new battery , ensure the battery is label side up. 5. P ositi on new battery on thermal pad as shown. 6. Connect the battery wirin g harness to the PC board. 7. Install inlet filter with tubi ng into holder .
SECTION 3—CHECKING O 2 PURITY XPO 2 ™ 14 Part No 1150748 SECTION 3—CHECKING O 2 PURITY NOTE: Although the XPO 2 offers pu lse dose only , the oxygen purity can be checked with a hand held oxygen analyzer . NOTE: Oxygen purity should be checked every 4,320 hou rs.
SECTION 4—ERROR CODES Part No 1150748 15 XPO 2 ™ SECTION 4 —ERROR CODES The XPO 2 has dozens of alarms that are being monitored on a continuous basi s. The alarms are grouped into four main categories: • Common Alarms: These are alarms for whi ch the patient, or the provider , can take action to eliminate the is sue.
SECTION 4—ERROR CODES XPO 2 ™ 16 Part No 1150748 (*) Wi thin the System Alarms there are a few alarms that may co rrect themselves simply by cha nging the in ter nal batte ry .
NOTES Part No 1150748 17 XPO 2 ™ NOT ES.
NOTES XPO 2 ™ 18 Part No 1150748 NOT ES.
LIMITED WARRANTY Part No 1150748 19 XPO 2 ™ LIMITED WARRANTY For warranty informati on, please refer to th e origina l owner's manual which came with this produ ct, or contact Invacare f or more info rmation.
Invacare Corporation www.i nvacare. com USA One Invaca re Way Elyr ia, Oh io USA 44036-2125 800-333-6900 Canada 570 Matheso n Blvd E Unit 8 Missi ssauga Ont ario L4Z 4G4 Ca nada 800-668-5324 All rights re served. Trade marks are identi fied by ™ and ®.
デバイスShop-Vac XPO100の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Shop-Vac XPO100をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはShop-Vac XPO100の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Shop-Vac XPO100の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Shop-Vac XPO100で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Shop-Vac XPO100を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はShop-Vac XPO100の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Shop-Vac XPO100に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちShop-Vac XPO100デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。