Instruction Manual P6131 10X Passive Pr obe for 2400 Series Oscilloscopes 070-5514-03 *P070551403* 070551403 W arning The servicing instructions are for use by qualified personnel only . T o avoid personal injury , do not perform any servicing unless you are qualified to do so.
Copyright © T ektronix, Inc. All rights reserved. T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved.
W ARRANTY T ektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of one (1) year from the date of purchase from an authorized T ektronix distributor .
Contacting T ektronix Product support For questions about using T ektronix measurement products, call toll free in North America: 1-800-833–9200 6:00 a.m. – 5:00 p.m. Pacific time Or contact us by e-mail: tm_app_supp@tek.com For product support outside of North America, contact your local T ektronix distributor or sales office.
P6131 Instruction Manual i General Safety Summary Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. T o avoid potential hazards, use this product only as specified. Only qualified personnel should perform service pr ocedur es.
ii P6131 Instruction Manual Service Safety Summary Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service procedures.
P6131 Instruction Manual 1 Operating Basics The P6131 probe is a subminiature, 10X, passive probe designed specifically for use with T ektronix 2400 Series oscilloscopes. It is fully compatible with the T ektronix family of subminiature probe accessories.
Operating Basics 2 P6131 Instruction Manual Grounding the Probe Connect the probe to the instrument and connect the ground lead to ground before making any measurement. Ensure that no part of the ground lead contacts voltage in the circuit under test.
P6131 Instruction Manual 3 Specifications The characteristics listed in T able 1 apply to P6131 probes installed on T ektronix 2445 and 2465 oscilloscopes or 7A42 plug-ins, unless otherwise noted.
Specifications 4 P6131 Instruction Manual @25% Duty Factor =(1/2 Peak V) RMS @50% Duty Factor =(1.414 (Peak V)) RMS Special case of square wave: @20% Duty Factor =[(V Peak ) 2 (Duty Factor)] 1/2 Example: Maximum Input V oltage =[(650 V Pk) 2 (.
Specifications P6131 Instruction Manual 5 Frequency (MHz) 500 100 10 1 10 100 V oltage (RMS) Figure 4: T ypical P6131 voltage derating versus frequency T able 2: Environmental characteristics Characte.
Specifications 6 P6131 Instruction Manual T able 4: Certifications and compliances EC Declaration of Conformity Compliance was demonstrated to the following specification as listed in the Official Jou.
P6131 Instruction Manual 7 Service Information W ARNING. The following instructions ar e for use by qualified service personnel only . T o avoid electrical shock, do not disassemble or maintain the pr obe while it is connected to a signal sour ce other than those specified in this pr ocedur e.
Service Information 8 P6131 Instruction Manual T able 5: Performance verification and adjustment test equipment (cont.) Description Example product Minimum requirements T ermination 50 BNC T ektronix 01 1-0049-XX Low-reactance alignment tool 2 inch shaft, .
Service Information P6131 Instruction Manual 9 4. Adjust the sine wave generator output amplitude to produce a 6-division display on the test oscilloscope. Center the waveform on screen. 5. Set the sine wave generator to the high-frequency range and slowly increase the variable frequency control until the display amplitude decreases to 4.
Service Information 10 P6131 Instruction Manual 6. Remove the coax cable from the test setup. Connect the P6131 probe output to the same vertical input on the oscilloscope. 7. Connect the probe tip through the probe-tip-to-BNC adapter to the standard amplitude output of the calibration generator .
Service Information P6131 Instruction Manual 11 T o adjust the probe low-frequency compensation do the following: 1. Connect the probe output to the test oscilloscope vertical input. 2. Set the oscilloscope controls as follows: V olts /Division 0.1 V (includes 10X probe attenuation) T ime/Division 1 ms Input Coupling DC Bandwidth Full 3.
Service Information 12 P6131 Instruction Manual Over compensated Under compensated Correct Figure 7: Low-frequency compensation, optimizing the waveform T o adjust the probe high-frequency compensation do the following: 1.
Service Information P6131 Instruction Manual 13 8. Connect the probe output to the oscilloscope vertical input. 9. Connect the probe tip through the probe-tip-to-BNC adapter and a 50 W termination to the positive-going fast-rise output of the calibration generator .
Service Information 14 P6131 Instruction Manual R2010 R2020 R2021 C1010 R2010 C1010 C2010 R2021 R2020 Plastic-shell version Metal-shell version Figure 8: High-frequency compensation adjustment locatio.
Service Information P6131 Instruction Manual 15 Maintenance Use the following procedures to clean and maintain the P6131 probe. T o remove accumulated dirt from the probe exterior , use a soft cloth dampened with a nonresidue cleaner , preferably isopropyl alcohol.
Service Information 16 P6131 Instruction Manual.
P6131 Instruction Manual 17 Replaceable Parts This section contains a list of the replaceable modules for the P6131 probe. Use this list to identify and order replacement parts. Parts Ordering Information Replacement parts are available through your local T ektronix field office or representative.
Replaceable Parts 18 P6131 Instruction Manual Using the Replaceable Parts List This section contains a list of the replaceable mechanical and electrical compo- nents for the P6131 probe. Use this list to identify and order replacement parts. The table below describes the content of each column of the parts list.
Replaceable Parts P6131 Instruction Manual 19 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 Figure 10: P6131 probe with standard accessories Replaceable parts: P6131 probe and standard accessories Fig. & index number T ektronix part number Serial no. effective Serial no.
Replaceable Parts 20 P6131 Instruction Manual Replaceable parts: P6131 probe and standard accessories (cont.) Fig. & index number Mfr . part number Mfr . code Name & description Qty Serial no. discont’d Serial no. effective T ektronix part number 10 –6 174–0972–00 8805 1 CABLE ASSY ,RF:39 OHM COAX,3.
Replaceable Parts P6131 Instruction Manual 21 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Figure 1 1: P6131 probe optional accessories Replaceable parts: P6131 probe optional accessories Fig. & index number T ektronix part number Serial no. effective Serial no.
Replaceable Parts 22 P6131 Instruction Manual Replaceable parts: P6131 probe optional accessories (cont.) Fig. & index number Mfr . part number Mfr . code Name & description Qty Serial no. discont’d Serial no. effective T ektronix part number 1 1–5 206–0364–00 8851 1 TIP ,PROBE:MICROCKT TEST ,0.
デバイスTektronix P6131の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Tektronix P6131をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはTektronix P6131の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Tektronix P6131の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Tektronix P6131で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Tektronix P6131を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はTektronix P6131の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Tektronix P6131に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちTektronix P6131デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。