CS510 OXYGUARD DISSOLVED OXYGEN PROBE INSTRUCTION MANUAL 1/00 COPYRIGHT (c) 2000 CAMPBELL SCIENTIFIC, INC..
This is a blank page..
The CS510-L Oxyguard Dissolved Oxygen Probe is manufactured by Point Four Systems Incorporated. Point Four’s conditions of sale and return policy apply to these items. Products may not be returned without prior authorization. To obtain an RMA number contact Point Four Systems, Inc.
This is a blank page..
1 CS510 OxyGuard Dissolved Oxygen Probe An accurate and reliable sensor is a critical element in any measurement system. The OxyGuard Stationary Probe meets these criteria for the measurement of dissolved oxygen.
CS510 OxyGuard Dissolved Oxygen Probe 2 2. Specifications Principle of Measurement: Membrane covered polarographic oxygen probe — self polarizing galvanic cell Output Si gnal: 5 mV per mg/ l, 2 mV p.
CS510 OxyGuard Dissolved Oxygen Probe 3 3. Service Kit Spare membr ane cap with fitted membran e. 50 ml electrolyte. Spare anode and O-ring. FIGURE 3-1. Exploded View of CS510-L 4. Optional Probe Accessories Agitator for s t agnant condi t ions . Probe Mou nting assem bly for waste water installation s.
CS510 OxyGuard Dissolved Oxygen Probe 4 FIGURE 5-1. Preventing Bio-fouling of the CS510-L The PT4 Agitator overcomes thes e problems . The devi ce is des i gned so t hat a soft brist le brus h sw eeps across t he probe membrane or se ns or ti p.
CS510 OxyGuard Dissolved Oxygen Probe 5 Optional Rep eat Cycle Timer fo r Agitators : Reciprocating action may be controlled by the opt ional sol id s t ate R epeat C ycle Ti mer. It s ends 12 V DC puls e s to the agitator coil. The ON-time is 0. 25 sec.
CS510 OxyGuard Dissolved Oxygen Probe 6 3) Determine the air temperature and barometric pressure. 4) Usi ng a cali brati on chart s uch as that provided i n the probe ’ s manual, determine the oxygen concentrati on of the air.
CS510 OxyGuard Dissolved Oxygen Probe 7 The following instructions in program table 2 would trigger the agitator as discussed in the agitator manual: *Table 2 Program 02: 2.0 Execution Interval (seconds) 1: Do (P86) 1: 41 Set Port 1 High ; Use port number as wired 2: Excitation with Delay (P22) 1: 1 Ex Channel 2: 0 Delay W/Ex (units = 0.
CS510 OxyGuard Dissolved Oxygen Probe 8 This is a blank page..
デバイスCampbell Manufacturing CS510の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Campbell Manufacturing CS510をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはCampbell Manufacturing CS510の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Campbell Manufacturing CS510の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Campbell Manufacturing CS510で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Campbell Manufacturing CS510を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はCampbell Manufacturing CS510の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Campbell Manufacturing CS510に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちCampbell Manufacturing CS510デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。