DB1 Liquid Level Measurement Sensor Revision: 12/07 Copyright © 1993-2007 Campbell Scientific, Inc..
Warranty and Assistance The DB1 LIQUID LEVEL MEASUREMENT SENSOR is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in m aterials and workmanship under norm al use and service for twelve (12) m onths from date of shipment unl ess specified otherwise.
DB1 Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for links to specific sections. 1. Overview ...............................................................
This is a blank page..
DB1 Liquid Level Measurement Sensor 1. Overview The double bubbler is a self-cali brating system for l iquid level m easurement. This manual focuses on water level measurement in stilling wells o r open channels, but the inform ation applies to other l iquids as well.
DB1 Liquid Level Measurement Sensor 2. Specifications Measurement Range: Option - 5 (0 to 5 psi/0 to 11.5 ft.) Option - 15 (0 to 15 psi/0 to 34.5 ft.) Option - 30 (0 to 30 psi/0 to 69 ft .) Maximum Overpressure Range: Option - 5 20 psi Option - 10 45 psi Option - 15 60 psi Transducer Accuracy: ± 0.
DB1 Liquid Level Measurement Sensor fluctuates between 5 and 15 ft (2.2 a nd 6.5 psi ), set the pressure regulator above 7 psi (read 10 psi) to ensure adequate pressure. Open the nitrogen bottle valve, set the p ressure regulator, an d adjust the nitrogen flow by adjusting t he two needle valves.
DB1 Liquid Level Measurement Sensor FIGURES 1 and 2. The DB1 System 4.
DB1 Liquid Level Measurement Sensor CR10(X) , CR1000, CR 800 CR850, 21X FIGURE 3. Wiring Diagram for DB1 5.
DB1 Liquid Level Measurement Sensor 5. Programming Every 60 seconds, the following program examples calcul ate the pressure transducer multiplier, measure the offset, and measure the liquid level. The average liquid level i s stored in final m emory every 15 m inutes.
DB1 Liquid Level Measurement Sensor C a s e 3 P o r t S e t ( 3 , 1 ) 'Open valve to atmosphere EndSelect Delay(0,150,MSEC) S u m = 0 For j = 1 To 25 'Take 25 pressure readings for each tube.
DB1 Liquid Level Measurement Sensor 3: Excitation with Delay (P22 ) 1: 1 EX Chan 2: 0 Delay w/EX (units=.01sec) 3: 150 Delay after EX (unit s=.01sec) 4: 0 mV Excit ation Subroutine 1 measures the pres.
DB1 Liquid Level Measurement Sensor 2: Z=F (P30) 1: 0 F 2: 0 Exponent of 10 3: 7 Z Loc [:SUM ] 3: Beginning of Loop (P87) 1: 0 Delay 2: 25 Loop Count 4: Full Bridge (P6) 1: 1 Rep 2: 23 25 mV 60 Hz rej.
DB1 Liquid Level Measurement Sensor 13: Do (P86) 1: 1 Call Subrouti ne 1 14: Z=X (P31) 1: 8 X Loc AVG LEVEL 2: 9-- Z Loc [:m V/V #1 ] 15: Do (P86) 1: 51-- Set low Port 1 16: End (P95) 17: Z=X-Y (P35) .
DB1 Liquid Level Measurement Sensor 6.1 Periodic Maintenance • Inspect submerged bubbler t ubes for sediment or ot her debris. Remove debris and clean tubes as required. • Replace submerged bubbler tubes if cracked or dama ged. • Replace nitrogen bottle every 2 to 3 months dependi ng on bubble rate.
DB1 Liquid Level Measurement Sensor 12 7. Replacement Parts Part # Description Qty. in DB1 5199 SWAK THREAD SEALANT 1 7397 TUBE FITTING ELL 1/4 O.D.x 1/8NPT 4 7558 NEEDLE VALVE ASSEMBLY 2 7561 2-WAY NC SOLENOID VALVE 3 7571 PLUG PIPE 1/8NPT TEFCOAT 2 6078 MOUNTING SCREW (6/32 x 2.
This is a bl ank page ..
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 Nort h Logan, Utah 84 321 UNITED STATES www.campbellsci.com info@campbellsci.com Campbell Scientific Afri ca Pty. Lt d. (CSAf) PO Box 2450 Somerset West 7129 SOUTH AFRICA www.
デバイスCampbell Manufacturing DB1の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Campbell Manufacturing DB1をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはCampbell Manufacturing DB1の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Campbell Manufacturing DB1の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Campbell Manufacturing DB1で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Campbell Manufacturing DB1を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はCampbell Manufacturing DB1の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Campbell Manufacturing DB1に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちCampbell Manufacturing DB1デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。