Manufacturer OMR ON HEAL THCARE Co ., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAP AN EU-representativ e OMRON HEAL THCARE EUR OPE B.V . Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS www .
c d 2 1 L F a b J G H I J K L M N O P T Q R U V W X Y Z S A B C D E F Thank you for purchasing the OMRON M3 Digi tal Automatic Blood Pressur e Monitor . The OMRON M3 is a compact, fully automa tic blood pressure monitor , operating on th e oscillometric principle.
1 EN Import ant Safety Information Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis. Please read this section carefully before using the unit. Wa rn in g : • Indicates a potentially hazardous situation which, if not avoided, could result in d eath or serious injury .
2 Important Safety Information (AC Adapter Usage) • Use only the original AC adapter (optional) designed for this unit . Use of unsupported adapters may damage and/or may be hazardous to the unit. • Plug the AC adapter into the appropriate voltage outlet.
3 EN 1. Overview Main unit Display O/I ST ART ( ) button Memory button Up/Down ( ) buttons Buzzer and Date/T ime setting ( ) button Air Jack Battery compartment AC adapter jack (for optional AC adapte.
4 2. Prep aration 2.1 Inst alling/Replacing the Batteries 1. Remove the battery cover . 2. Insert four “AA” batteries as indicated in the battery comp artment and then re place the battery cover . Notes: • If t he battery low symbol ( ) appears on th e display , turn th e unit off then replace all bat teries at the same time.
5 2. Preparation EN 2.2 Setting the Buzzer/ Date and T ime 1. Press the button. 2. Set the Buzzer . 3. Set the monitor to the corre ct date and time before t aking a measurement for the first time.
6 3. Using the Unit 3.1 Applying the Arm Cuff Remove tight-fitting clothing or tight rolled up sleeve from your up per arm. Do not plac e the cuff over thick clot hes. 1. Insert the air plug into the air jack . 2. Put your arm through the cuff loop. 3.
7 3. Using the Unit EN 3.2 How to Sit Correctly T o take a measurement, you ne ed to be relaxed and comfort ably seated, under comfortabl e room temperature. No eating , smoking or exercising 30 minutes before taking a m easurement. • Sit on a chair with your feet flat on the floor .
8 3. Using the Unit 3.3 T aking a Reading Notes: • T o cancel a measurement, press the O/I ST ART button to turn off the unit and to release the air in the arm cuf f . • Remai n still while taking a measurement. 1. Press the O/I ST ART button. The cuff will start to inflate automatically .
9 3. Using the Unit EN 2. Remove the arm cuff. 3. Press the O/I ST ART button to turn off the monitor . The monitor auto matically stores the measurement in its memory .
10 3. Using the Unit 3.4 Using the Memory Function The monitor automatically stores the result up to 60 sets. It can also calcula te an average reading base d on the measurement s from the last three readings t aken within 10 minutes. If there are only two reading s in memory for that period, the average will be based on two readings.
11 3. Using the Unit EN T o delete all the val ues stored in memory When the memory symbol ( ) appears, firs t press the Memory button. Then while holding it down, press the O/I ST ART button simult aneously for about 2-3 seconds. Note: Y ou cannot partially delete the stored reading s.
12 4. T roubleshooting and Maintenance 4.1 The Icons and Error Messages Error Display Cause Remedy Irregular or weak pulses are detected. Remove the arm cuff. W ait 2-3 minutes and then take another measurement. Repeat the steps in section 3.3. If this error continues to appear , contact your doctor .
13 4. Troubleshooting and Maintenance EN Note: The irregular heartbe at symbol ( ) may also be displayed with error messages. Cuff is under inflated. Carefully read and repeat the steps listed under section 3.3. Movement during measurement. Repeat measurement.
14 4. Troubleshooting and Maintenance 4.2 T r oubleshooting Problem Cause Remedy The reading is extremely low (or high). Arm cuff not applied correctly . Apply the arm cuff correctly . Refer to section 3.1. Movement or talking during measurement. Remain still and do not talk during measurement.
15 4. Troubleshooting and Maintenance EN Nothing happens when you press the buttons. The batteries are empty . Replace the batteries with new ones. Refer to section 2.1. The batteries have been inser ted incorrectly . Insert the batteries with the correct (+/-) polarity .
16 4. Troubleshooting and Maintenance 4.3 Maintenance T o protect your unit from damage, please ob serve the following: • Do not subject the main unit and the cuff to extreme te mperatures, humidity , moisture or direct sunlight. • Do not fold the cuff or tubing tightly .
17 4. Troubleshooting and Maintenance EN 4.4 St or a ge Keep the unit in it s storage case when not in use. 1. Unplug the air tube fr om the air connector . 2. Gently fold the air tu be into the arm cuff. Note: Do not bend the air tube excessively . 3.
18 5. Optional Part s Using the Optional AC Adapter 1. Insert the AC adapter plug into the AC adapter jack on the rear side of the main unit. 2. Plug the AC adapter into an electrical outlet. T o disc onnect the AC adap ter , unplug the AC adapte r from the electrical outlet first and then remove the AC ad apter plug from the main unit.
19 EN 6. T echnical Dat a Note: Subject to technical modi fication without prior notice. Product Description Digital Automatic Bloo d Pressure Monitor Model OMRON M3 (HEM-7200-E(V )/ HEM-7200-E8(V)) Display LCD Digital Display Measurement Method Oscillometric method Measurement Range Pressure: 0 mmHg t o 299 mmHg Pulse: 40 to 180/min.
20 6. Technical Data • This device fulfils the provisions of EC di rective 93/42/EEC (Medical Device Directive). • This blood pressure monitor is designed according to the European S tandard EN106.
21 6. Technical Data EN Import ant informatio n regarding Elec tro Magnetic Comp atibility (EMC) With the increased number of electronic devices such as PC’s and mobile (ce llular) telephones, medical d evices in use may be susceptible to electromagnetic interference fr om other devices.
22 7. Some Useful Information about Blood Pressure What is Blood Pressure? Blood press ure is a measure of the force of bl ood flowing against the walls of the arteries. Arterial blood pressure is co nstantly changing during the course of the heart’s cycle.
23 7. Some Useful Inform ation about Blood Pressure EN Why is it a Good Thing to measure Blood Pressure at Home? Many factors such as physical activity , anxiety , or the time of day , can influence your blood pressure . A single measurem ent may not be s uf f icient for an accurate diagnosis.
24 7. Some Useful Inform ation about Blood Pressure Classification of Blood Pressure by the W orld Health Organization The World Health Organiz ation (WHO) and the International Society of Hypertension (ISH) developed the Blood Pressure Classificat ion shown in this figure.
デバイスOmron M3の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Omron M3をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはOmron M3の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Omron M3の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Omron M3で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Omron M3を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はOmron M3の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Omron M3に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちOmron M3デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。